Plate-reader assays

After reading a plate, most labs do one of a few standard things: subtract blanks and average replicates, fit a standard curve and back-calculate unknowns, fit dose-response curves, extract kinetic or growth rates, or check the plate’s quality from its controls. openreadout analyze has one subcommand for each, with a documented method.

Here is a dose-response plate with three compounds, positive and negative controls, and a layout file that says which well holds what:

Terminal window
openreadout analyze dose-response plate.csv --layout layout.csv
plate.csv — dose-response of read 1 `RLU`
layout: layout.csv (negative 8, positive 8, sample 80)
outliers (grubbs): F10 (flagged, kept)
CPD-A: IC50 0.235142 (95% CI 0.199935–0.27655), Hill -0.964425, top 100988.558518, bottom 1952.059038
CPD-B: IC50 1.825145 (95% CI 1.566927–2.125916), Hill -1.558539, top 99762.40999, bottom 2172.856099
CPD-C: IC50 4.28051 (95% CI 0.460937–39.751089), Hill -0.695743, top 100173.006341, bottom 9587.948892
...

The files are in the repository as crates/openreadout-assay/tests/fixtures/assay-synth-dose-response*.csv.

Commands

Terminal window
openreadout analyze assay-wells FILE [--layout L.csv] [--blank-wells H1,H2] # blanks, replicates, CVs, outliers
openreadout analyze assay-curve FILE [--layout L.csv] [--standard A1,A2=100] [--model linear|4pl|5pl]
openreadout analyze dose-response FILE --layout L.csv [--normalize controls] [--model 4pl|5pl]
openreadout analyze kinetics FILE [--window N]
openreadout analyze growth FILE [--window N]
openreadout analyze assay-qc FILE [--positive-wells A1:H1 --negative-wells A12:H12]

Options shared by every subcommand:

  • --json prints the result in the JSON wrapper. The fields are listed in assay.
  • --csv PREFIX writes tidy tables: PREFIX.wells.csv, .samples.csv, .compounds.csv, .kinetics.csv, .growth.csv and .curve.csv.
  • --table N picks the plate in a file with several.
  • --read R picks the read by number or label. The default is the first measured read; for kinetics and growth, the first kinetic read.
  • --wavelength-nm NM picks a wavelength of a spectral read.
  • --plot FILE.png draws the fitted curve (assay-curve and dose-response).

Over MCP the six subcommands are one tool, openreadout_assay, with analysis set to wells, curve, dose-response, kinetics, growth or qc and the subcommand’s arguments. The tool takes layout and the control wells at the top level, and the rarely needed options (table, read, wavelength_nm, blank_subtraction, outliers, roles, …) in a plate_options object. In Python they are openreadout.analyze(path, "assay", analysis=…, layout=…, **options). To run them over many plates, see batch.

Input. Any plate-reader export OpenReadout reads (see Plate readers), or a long CSV or TSV file. A long file has a well column and a value column named value, signal, od, absorbance, fluorescence, luminescence, rfu, rlu or similar. For kinetics it also has a time column (time_s, time_min, time_h or time). It may carry read, wavelength_nm and layout columns such as role and sample.

Layouts

A layout says what each well holds. Names are matched without regard to case, and common synonyms are accepted.

  • role (or type, well_type): one of
    • blank (also buffer, background, no-cell, medium only)
    • standard (also std, calibrator, cal)
    • sample (also unknown, unk, test)
    • positive (also pos, pc, high control, max)
    • negative (also neg, nc, low control, min)
    • control, a control whose sign is not known (also ctrl, qc, vehicle, DMSO, untreated)
    • empty (also none, unused)
  • sample (or sample_id, name, label): the replicate-group name. When there is no role, the role is read from the name: STD3 and S1 are standards, BLK is a blank, POS1 and PosCtrl are positive controls, NEG1 and NegCtrl are negative controls, DMSO and vehicle are controls, and anything else is a sample.
  • compound (or drug, treatment, inhibitor): the dose-response series.
  • concentration (or conc, dose): a number. A unit after it (1 microg/ml) or in the header (concentration (nM)) is reported as layout.concentration_unit.
  • dilution (or dilution_factor, df): 1:10, 1/10, 10x and 10 all mean a factor of 10. A plain number below 1 is a fraction, so 0.5 means 2.
  • group (or replicate): an explicit replicate group.
  • Any other column is kept under extra.

A layout file can have either of two shapes, separated by commas, tabs or semicolons. These are the same shapes that sample sheets accept.

  • Plate-map grid: one block per field. The header line has the field name in its first cell and the column numbers 1 2 3 …; then one line per row letter. Blank lines separate blocks.
  • Long table: a header with a well column (A1 or A01, or row and column) and one line per well.
role,1,2,3,4
A,blank,standard,standard,sample
B,blank,standard,standard,sample
concentration,1,2,3,4
A,,100,50,
B,,100,50,

Layout information is merged well by well, and later sources override earlier ones:

  1. The layout the export embeds, such as Gen5 Well ID and Conc/Dil, SkanIt sample matrices and layout sheets, or BMG content maps. --no-embedded-layout ignores it.
  2. --layout FILE.
  3. The well flags --blank-wells, --positive-wells, --negative-wells and --empty-wells (wells as H1,H2, A1:H1 or A1-A12), and --standard WELLS=CONC (repeatable).

Roles

Some names do not say which way a control points. A vehicle or untreated well is the no-effect control of an inhibition assay but the full-signal control of an activation assay. Such wells are controls of unknown sign and are left out of normalization and Z′ until you map them:

Terminal window
openreadout analyze dose-response plate.xlsx --layout layout.csv --role DMSO=negative

--role NAME=ROLE (repeatable) sets the role of every well whose role text or sample name is NAME. --role CTL=positive also covers CTL1 and CTL2. It overrides every other source.

The output’s warnings list what may be wrong with the layout: a role text that is not a role name, controls of unknown sign, and --role entries that matched no well. Each warning names the --role argument that fixes it.

Steps

  1. Read the selected read and wavelength. A kinetic read used in an endpoint analysis (wells, curve, dose-response, qc) needs --reduce:

    • max-slope: the largest slope over a sliding window, per minute
    • mean-slope: the least-squares slope over all points, per minute
    • first, last, max, min, mean
    • auc: the trapezoidal area, value × seconds
  2. Group replicates by the explicit group, else the sample name, else compound and concentration, else role.

  3. Flag outliers on the values as read, one per group at most:

    • --outliers grubbs (the default): the two-sided Grubbs test at alpha 0.05, in groups of at least 3 wells. --outlier-threshold changes alpha.
    • --outliers mad: a modified z-score above 3.5, in groups of at least 5 wells.
    • --outliers none.

    Flagged wells are kept unless --exclude-outliers is given.

  4. Subtract blanks. --blank-subtraction auto (the default) subtracts the mean of the blank wells when there are any. mean and median require blanks; none skips the step. Kinetic analyses subtract blanks per time point.

  5. Summarize each group: n, n_used, mean, sd (n − 1) and cv_percent.

  6. Run the analysis.

Standard curves

analyze assay-curve fits the standard wells and back-calculates the concentration of every other well.

model formula
linear y = slope · x + intercept
4pl (default) y = d + (a − d) / (1 + (x/c)^b)
5pl y = d + (a − d) / (1 + (x/c)^b)^g

In the 4PL, a is the response at zero concentration, d the response at infinite concentration, c the inflection point and b > 0 the slope factor. This is the 4PL of Gen5 and SkanIt. In the 5PL, g > 0 is the asymmetry.

Fitting. The fit minimizes the weighted sum of squared residuals. --weighting is none (the default), 1/y, 1/y2, 1/x or 1/x2. Logistic models are fitted from several starting points, and the best fit is kept. A concentration of 0 is allowed. The fit needs at least 2 distinct concentrations for linear, 4 for 4pl and 5 for 5pl. --fit-on replicates (the default) fits every standard well; --fit-on means fits the mean of each level.

Fit results (curve.fit):

  • parameters[], each with value, se and a confidence interval (--confidence, default 0.95). Standard errors follow the convention of SciPy curve_fit, R nls and GraphPad.
  • r_squared, residual_se, sse, df, n_points, n_levels, converged
  • points[], with each point’s x, y, fitted and residual

Back-calculation. Every well and every group mean is converted through the inverse of the curve. flag says how far to trust it:

  • in_range: within the standards’ concentration range
  • below_range, above_range: outside it; the extrapolated concentration is still reported
  • below_curve, above_curve: the signal is beyond an asymptote, so there is no concentration
  • not_computable: the curve is flat

final_concentration is the back-calculated value times the dilution. Standards also get recovery_percent. LLOQ and ULOQ are the lowest and highest standards whose back-calculated mean is within 20 % of nominal with a CV of at most 20 %; --lloq and --uloq set them yourself. quantifiable says whether a well lies between them.

Dose-response

analyze dose-response fits a 4PL (default) or 5PL curve per compound to the sample wells that have a concentration.

--normalize controls first converts every value to percent effect: 100 × (y − negative mean) / (positive mean − negative mean). The negative control is 0 %, the positive control 100 %. On analyze assay-wells, the same option adds percent_effect to every well.

Each entry of compounds[] has:

  • ec50 and kind: IC50 when the response falls with dose, else EC50
  • ec50_ci_low, ec50_ci_high: the confidence interval, computed on the log scale so it is always positive
  • log10_ec50, and absolute_ec50 when normalized (where the curve crosses 50 %)
  • hill_slope: positive when the response rises with dose, as in GraphPad
  • top, bottom, response_at_zero, response_at_infinity
  • min_concentration, max_concentration, and extrapolated when the EC50 lies outside them
  • error, for example when there are fewer than 4 doses or the fit did not converge

Kinetics

analyze kinetics computes per well of a kinetic read. --window sets the points per sliding window; the default is 5 or a tenth of the time points, whichever is larger.

field meaning
max_slope_per_min the largest least-squares slope over any window (Vmax), per minute
max_slope_r_squared, window_start_s, window_end_s that window’s fit and limits
lag_time_s where the max-slope line crosses the baseline
mean_slope_per_min the least-squares slope over all points
max_value, time_to_max_s, initial_value, final_value as named
auc trapezoidal area under the values, value × seconds

Growth curves

analyze growth computes per well after background correction. The background is the blank wells per time point when there are blanks, else the well’s own minimum, as in growthcurver. Blank wells get no row of their own. --wells (MCP wells) narrows the wells reported, and the layout’s blank wells outside it are still subtracted; a note names them.

field meaning
growth_rate_per_h µmax: the largest slope of ln(value) against time in hours over any window, using points above --threshold (default 5 % of the well’s maximum)
doubling_time_h ln 2 / µmax
exp_phase_start_s, exp_phase_end_s that window
lag_time_s where the µmax line crosses the starting level
logistic_k, logistic_r_per_h, logistic_n0, logistic_doubling_time_h a least-squares fit of the logistic curve N(t) = K / (1 + ((K − N0)/N0) e^(−rt)), the model growthcurver fits; only for wells that grew, since a crash breaks the logistic shape
max_value, time_to_max_s, auc_h as named; area in value × hours
pattern grew, no_growth, crashed (fell to half its height above its lowest value, by at least 5 noise SDs, and stayed down) or regrew (fell, then rose again by half the fall and 10 noise SDs). Measured on a running median, so single noisy reads do not count. Growth and regrowth must also rise by a tenth of the larger rises on the plate (a note gives the value), and a fall with no growth before or after it is no_growth, so a flat well’s noise is not read as a crash
crash_s, crash_value the high point before the fall: the decline starts there (0 when the values fall slowly from the first read, as when phage lyse the inoculum). A decline that ends within the first hour, such as condensation clearing or cells settling, is not a crash: a crash after it needs a rise of 5 noise SDs first
trough_s, trough_value the lowest point after the fall
regrowth_s when growth restarts: the lag time of the part after the trough
exp_phase_after_crash the exponential phase above is the regrowth, so the rate and doubling time are those of the cells that came back

growth_summary lists the wells of each pattern and the smallest, median and largest crash and regrowth times, so “how many infected wells came back, and when” is one call.

Over MCP, openreadout_assay rows keep the main fields above (rate, doubling time in hours, lag, maximum, pattern and the crash and regrowth times) to 4 significant digits, so a whole 96-well plate fits in one reply; fields: ["all"] returns every field. See Per-well rows.

The two rates answer different questions. µmax is the steepest exponential growth seen. The logistic r belongs to the best-fitting logistic curve and is smaller when the growth is not logistic. Say which one you report.

Assay quality

analyze assay-qc reports plate quality from the control wells. The other analyses add the same quality block whenever the layout has positive and negative controls. It uses the values before blank subtraction.

  • positive, negative, blank and samples: group statistics
  • z_prime: 1 − 3 (SD₊ + SD₋) / |mean₊ − mean₋| (Zhang, Chung and Oldenburg, 1999), with assessment excellent (at least 0.5), marginal (0 to 0.5) or unusable (0 or below)
  • z_factor: samples against the negative control
  • signal_to_background, signal_to_noise and ssmd
  • median_replicate_cv_percent, over groups of two or more wells

Errors

Mistakes in the request are usage errors (exit 2) with a message that says what to add: no standards or concentrations, no controls for assay-qc or --normalize controls, a kinetic read without --reduce, a spectral read without --wavelength-nm, an endpoint read for kinetics or growth, too few concentrations for the model, or a bad layout line. A file that is not a plate gives the same exit code as info would.

Validation

The curve fits, back-calculation, dose-response, kinetics, growth and quality metrics are compared with independent implementations and with the results that plate-reader software stores in its exports. How we validate explains the method, and the reference tools and data are logged in docs/provenance/plate-analysis.md.