JASCO Spectra Manager .jws

JASCO Spectra Manager saves FT-IR, Raman, UV-Vis, circular dichroism and fluorescence measurements as .jws files. OpenReadout returns every channel as a spectrum with its axis, plus instrument, sample and acquisition settings (table below). Derived from public files written by FT/IR-4600, FT/IR-4700, NRS-5100, V-630, V-730, J-1500 and FP-8300 instruments, Spectra Manager text and CSV exports of the same measurements, and jws2txt (MIT) as a second opinion. Provenance: docs/provenance/jasco-jws.md. Crate: openreadout-spectro.

format id files reads confidence
jasco-jws .jws of JASCO Spectra Manager (compound-file and flat containers); .jrs from the instruments’ firmware (flat) every channel as a spectrum (%T, %R, absorbance, single-beam, intensity, CD, HT voltage), its axis (wavenumber, Raman shift, wavelength, time), instrument model and serial, sample name and comment, measurement time, FT-IR acquisition settings high (evidence rubric, docs/assurance.md)

Not read: interval/kinetics files (.jwb), spectra-manager projects, JASCO’s text exports (they are text; jcamp-dx-like but not JCAMP).

Containers

The first bytes decide; the extension is the same.

Compound file (D0 CF 11 E0; Spectra Manager 1.x and 2.x)

stream content our reading
Header UTF-16 L~, SPCMAN2, R2.00.00, 80x86, VC++6.0 detection (L~ … SPCMAN); format_version
DataInfo u32 version (3); u32; u32; u32 channel count; u32 grid flag (1 regular, 0 explicit X-Data); u32 points; f64 first x, last x, step (negative when descending); four u32 descriptors: the x axis, then one per channel, the remaining slots repeating from the start; then the saved plot view. 96 bytes for one channel, +44 per further channel axis, channels
Y-Data float32, channel-major, exactly 4 × channels × points bytes one trace per channel
X-Data float32 per point (when the grid flag is 0: Raman CCD axes) irregular axis (trace channel 0)
ModuleInfo u32 version, module name, u16, u16 module id, model, serial (UTF-16 strings with a u32 byte length) instrument
SampleInfo u32, sample name, comment, a record count, then records (u32 tag, u16 type, value); record 1 is the measurement time (OLE date, UTC) sample, comment, started_at
MeasParam u32, u32, a record count, records FT-IR settings (module 9, below); other instruments’ records raw in the vendor tree
BaseInfo u32, 16-byte id, u8, original path, f64 saved, f64 measured (OLE dates, UTC) vendor tree; started_at when SampleInfo/MeasParam have none (Raman)

Record value types: 2 u16, 3 u32, 4 f32, 5 and 7 f64, 8 UTF-16 string; records after a type not in this list are kept as an undecoded byte count.

Axis descriptors: 0x10000100 wavenumber (1/cm, INFRARED SPECTRUM), 0x10000101 Raman shift (1/cm, RAMAN SPECTRUM), 0x10000103 wavelength (nm), 0x20000203 time (a V-630 time course; no unit: the time unit is not in a validated field). Any other descriptor is refused (exit 6).

Channel codes: 0x00 transmittance (%), 0x03 absorbance (AU), 0x08 single-beam (FT-IR background), 0x0E intensity (FP-8300 fluorescence, NRS-5100 CCD counts), 0x1001 circular dichroism (mdeg when the file’s CD sensitivity text, MeasParam record 20, says mdeg), 0x2001 HT voltage (V). A code not in this list is returned as channel_0x… with a finding.

Refused (exit 6, with a hint): DataInfo version other than 3, more than three channels, descriptor slots that do not follow the pattern or hold an axis descriptor where a channel code belongs, grid flags other than 0 and 1. Y-Data/X-Data sizes that do not match are corrupt (exit 4).

Flat file (L~S ; Spectra Manager 2.x, instrument firmware)

offset content
0x08 SPECMAN (Spectra Manager) or SPECIRM (firmware)
0x20 R2.0.0
0x84 i32 points
0x88, 0x90, 0x98 f64 first x, last x, step
0x82 u16 channel count: 1, or 2 in circular-dichroism files
0xA0 x unit: 0 cm⁻¹, 3 nm (followed by 01 00 10: the compound files’ x descriptor)
0xA4 one u32 channel code per channel, the compound files’ codes: 0 %T, 2 %R, 3 absorbance, 9 single-beam reference, 10 single-beam sample; 0x1001 CD then 0x2001 HT voltage in a CD file
0xC8 i64 data length (= 4 × points × channels)
0x140, 0x160, 0x180, 0x1C0 model, serial, title, comment (NUL-terminated)
0x2C0 i32 time (Unix seconds, UTC)
0x740 to the end float32 y values, channel by channel (the data run from file length − data length, which is 0x740 in every file; any other value is refused as a truncated or extended file)

Other container ids, versions and x units are refused, and so are channel counts and codes other than one %T, %R, absorbance or single-beam channel, or CD with HT (the refusal names the count and the codes). The flat header does not say that CD values are in mdeg, so the CD unit is left unset.

What the reader returns

  • One trace per channel, one sweep: the channel’s name as trace name and y quantity (transmittance, absorbance, reflectance, single_beam, single_beam_reference, single_beam_sample, intensity, circular_dichroism, ht_voltage), values as stored (float32; JASCO’s −1.18e-38 invalid-point marker is kept, as its exports print it). extra.axis = {quantity, unit, first, last, step, size} with x = first + step × i (the stored last x is only compared: a finding when it disagrees by more than one step); an X-Data axis is {irregular: true, channel: 0} and the trace has two channels. extra.data_type: INFRARED SPECTRUM, RAMAN SPECTRUM, UV/VIS SPECTRUM, CIRCULAR DICHROISM SPECTRUM, FLUORESCENCE SPECTRUM, UV/VIS KINETICS; extra.channel_code, extra.title.
  • Experiment: instrument vendor JASCO, model and serial; sample name; acquisition comment and started_at (UTC); for FT-IR (module 9) method parameters accumulations, resolution (cm⁻¹), aperture (mm), scan_speed (mm/s), gain, filter (Hz), light_source, detector — named from the exports’ footers (records 1, 2, 3, 4, 6, 32, 47, 48).
  • Vendor tree jasco: the container, header text, DataInfo fields and descriptors, module, sample records, measurement-parameter records (tag, type, value), base info (original path, saved and measured times), the stream list; for flat files every header field.
  • info --view structure: the compound file’s streams, or the flat file’s header and data blocks.

Validation

cargo test -p openreadout-corpus-tests --features corpus (tests/jasco_oracle/mod.rs, oracle/jasco_oracle.py):

  • Spectra Manager exports of five measurements (FT-IR %T compound files in two locales, the Raman compound file with its X-Data axis, the V-730 flat file, the J-810 flat CD file with its HT column): point count, axis unit, y quantity, x within 0.005 and y to the export’s six significant digits at 256 rows; model and serial; the eight FT-IR settings; the measurement time equal to the export’s local time minus whole hours. One more export has the same name but is another acquisition: its instrument facts only.
  • jws2txt (MIT, black box) on every compound file: the same channels (count, names), point counts and values at 64 points per channel (CD/HT/absorbance files included).
  • One further depositor’s file is held out (docs/benchmark/heldout.md).

Vocabulary (public API of openreadout-spectro, JASCO)

identifier meaning
JwsReader reader of JASCO .jws/.jrs files
JWS_FORMAT_ID jasco-jws
SpectroDataset the dataset the spectroscopy readers share