Waters Empower keeps raw data and results inside its own database, so data leaves Empower only through exports. The ASCII raw-data export, saved with the .arw extension, holds one chromatogram as text. OpenReadout returns it as one trace with the sample, method and injection fields the export includes. The AIA/ANDI netCDF export (.cdf) is read by the andi-chrom reader. Sony .ARW camera files share the extension but are TIFF and never start with a quote.

Format id empower-arw. Code: crates/openreadout-chrom/src/empower_arw.rs. How each fact was established: docs/provenance/empower-arw.md.

Layout

Text (ANSI/UTF-8), rows separated by CR, LF or CRLF, cells by tabs:

  1. the names of the fields the export method includes, each in double quotes ("SampleName", "Channel", "Sample Set Name", "Instrument Method Name", …);
  2. their values, in double quotes, as many as names;
  3. one row per point: retention time (minutes) and the detector value, unquoted decimal numbers.

The value’s unit is not in the export (it is the detector’s: mV, AU, EU for fluorescence, …). Times are printed to 7 significant digits, so an evenly sampled run shows steps that differ in the last digit; points within 10⁻⁶ of the grid (relative to the time) are a regular trace.

Some exports write the header as one field per line instead, "name"<TAB>value, the value quoted text or an unquoted number, followed by the same data rows (GPCreader’s and HPLC-RS’s test exports). The reader takes the lines before the first row of two numbers as this layout when there are more than two of them, or when the first value is unquoted, and each is a quoted name and one value (extra.one_field_per_line true; assurance layout one field per line). Two quoted lines are read as a names row and a values row, the layout every two-line development export has.

An export method that selects no fields writes the data rows alone. Such a file is read when it has the .arw extension and its first lines are all two tab-separated numbers (extra.headerless true, no fields, the channel named value). Nothing else in it names Empower, so without the extension it is not claimed. No public export of this layout was found, so the assurance layout headerless export is unvalidated and --strict refuses its trace.

Detection: a first line of tab-separated, double-quoted cells (with or without the .arw extension); in a file named .arw, also a first line of a quoted name and a number, or rows of two numbers. Rows of more than two columns (a multi-wavelength PDA export) are refused (exit 6); a header whose value count differs from its name count, or a data row that is not two numbers, is corrupt (exit 4).

Mapping to the data model

  • One trace named <SampleName> / <Channel> (or the channel alone), one channel named after Channel without surrounding spaces (else value), dtype float64, no unit. Evenly spaced times: sample_rate_hz, start_s and extra.axis (retention time, minutes); otherwise sample_rate_hz 0 and a first channel time (minutes).
  • extra: fields (every exported name and value), headerless and one_field_per_line (only when true), and from them sample_name, channel, sample_set, instrument_method, processing_method, vial, injection, injection_volume, acquired_by, acquired_at when exported; line_ending, x_start_min, x_end_min, time_channel.
  • check: time_not_increasing (warning), irregular_times, no_channel_field (info).

Validation

  • Six exports of Appia’s test data (MIT, two fluorescence channels of three samples, 6,601 points each): every value equal to Appia’s own reading (processed-tests/…hplc-wide.csv, an independent reader) and the times within Appia’s single-precision rounding (2·10⁻⁶ min).
  • Ten GPC exports of GPCreader’s test data (MIT; four with a names row and a values row, six with one field per line) and seven HPLC exports of HPLC-RS’s test data (GPL-2.0, one field per line): every point equal to chromConverter’s reading (GPL-3.0, run as a black box).

Known gaps

  • Multi-column (3D PDA) exports: no licensed public example; refused.
  • Empower’s native data (database) and its report exports are not read; AIA/netCDF exports are read by andi-chrom.

Vocabulary (every public identifier in empower_*.rs must appear here)

identifier meaning
EmpowerArwReader, EmpowerArwDataset, EMPOWER_ARW_ID, open, export reader, opened export, the id empower-arw, open by path, the parsed export
MAX_ARW_BYTES largest export read (512 MiB)
ArwExport, fields, times, values, line_ending, headerless, one_field_per_line, field, regular_step a parsed export: header names and values, points, line ending, whether it has no header rows, whether its header has one field per line; a named field’s value; the step of an evenly spaced export
ArwError { Corrupt, Unsupported }, parse_arw, looks_like_arw, looks_like_arw_field_lines, looks_like_headerless_arw why an export was not read; the parser; detection of an export with quoted header rows, with one field per line starting with a number, and without header rows

How this reader was derived, file by file: provenance log.