Bruker flex MALDI-TOF (fid + acqus)

flexControl, the acquisition software of Bruker’s MALDI-TOF instruments (autoflex, ultraflex, microflex, MALDI Biotyper, rapifleX), writes each spectrum as a directory holding fid (the summed detector signal) and acqus (its parameters). OpenReadout reads every spectrum below the path it is given, with m/z from the time-of-flight calibration in acqus. Bruker publishes no description of these files. The reader follows OpenMS’s XMASS reader (BSD-3-Clause), read as documentation, and was checked against pyOpenMS run on the same files. Provenance: docs/provenance/bruker-flex.md. Reader: crate openreadout-bruker-tims, module flex, format id bruker-flex.

Layout

A run is a directory tree <run>/<spot>/<acquisition>/<kind>/, for example 0_A4/1/1SLin/:

  • <spot> names the target position (0_A4);
  • <acquisition> numbers the acquisitions of that spot (1, 2);
  • <kind> says how the spectrum was acquired: 1SLin linear, 1SRef or 1Ref reflector, and <precursor m/z>.LIFT/1SRef a LIFT fragment spectrum.

A spot directory (is_spot) holds fid and an acqus (or acqu) whose first lines name XMASS (##TITLE= XMASS Parameter file or ##ORIGIN= XMASS, Bruker-Daltonics). flexAnalysis adds pdata/1/ with the processed spectrum 1r and its procs. The reader does not read pdata.

Detection. The input is a spot directory, its fid or acqus, or any directory above spot directories. find_spots walks at most 7 levels down and skips pdata; detection looks at no more than 400 directories. A directory above spots counts only when each of its subdirectories holds a spot (hidden folders, __MACOSX and pdata aside), so a folder that also holds other data is not taken for a flex run. Bruker NMR (TopSpin) directories also hold fid and acqus, but their acqus does not name XMASS, so the NMR reader leaves flex directories alone and the flex reader leaves NMR directories alone.

Parameters

acqus is JCAMP-DX style text: ##$KEY= value lines (some keys use ##. or no prefix), strings in angle brackets, arrays as (0..n) followed by value lines. The reader keeps every key, with its first line of text for arrays. Keys it uses:

key meaning
TD number of points in fid
BYTORDA byte order of fid: 0 little-endian, 1 big-endian
DW, DELAY time per point and time of the first point (ns)
ML1, ML2, ML3 calibration constants (below)
IONIZATION MODE LD+ positive ions, LD- negative ions
AQ_mod, SPType acquisition mode: in the corpus AQ_mod 1 is linear, 2 reflector, and SPType 2 marks a LIFT spectrum
Parent the LIFT precursor m/z
PATH the spot directory as the instrument wrote it
NoSHOTS, INSTRUM, InstTyp, FCVer, AQ_DATE, CMT1, SPOTNO, PATCHNO laser shots, the instrument PC’s host name, a numeric instrument type, the flexControl version, the acquisition time, a comment, the spot name

acqus does not name the instrument model.

Points and m/z

fid is TD signed 32-bit integers in the BYTORDA byte order. Point i has the flight time tof = DW · i + DELAY (ns). With b = √(10¹² / ML1) and c = ML2 − tof, s = √(m/z) solves ML3 · s² + b · s + c = 0:

  • ML3 = 0: s = −c / b;
  • otherwise s = (√(b² − 4 · ML3 · c) − b) / (2 · ML3);

and m/z is s² (Calibration, mz). A spot whose acqus has no usable DW, DELAY, ML1 or ML2, or a non-positive ML1, is listed with a note and its points cannot be read.

LIFT spectra. A LIFT fragment spectrum gets m/z from the same constants. flexControl stores the fragment calibration in NTBCal, whose layout is not public, and the reader does not apply it, so fragment m/z can differ from flexAnalysis’s. notes and each such spectrum’s extra.mz_note say so.

Spectra

The run is one list of spectra, one per spot directory in path order. Each spectrum is a profile (centroided false), MS level 1, or 2 for LIFT with the precursor from Parent; polarity from IONIZATION MODE. native_id is the spot directory relative to the input. Intensities are 32-bit floats, exact up to 2²⁴.

key meaning
spectra[0].extra.modes spectra per acquisition mode (linear, reflector, LIFT, unknown)
spectra[0].extra.total_points, instruments, instrument_types, software_versions points in all spectra, INSTRUM, InstTyp and FCVer values
spectra[0].extra.spots the first 1000 spots: path, spot, mode, points
spectrum extra.spot, mode, points spot name (SPOTNO, else PATCHNO, else the path), acquisition mode, point count
spectrum extra.shots, instrument, instrument_type, software_version, acquired, comment NoSHOTS, INSTRUM, InstTyp, FCVer, AQ_DATE, CMT1
spectrum extra.calibration DW, DELAY, ML1, ML2, ML3
spectrum extra.processed pdata/1/1r when flexAnalysis wrote a processed spectrum
spectrum extra.mz_note LIFT spectra: the fragment calibration is not applied

The acquisition mode comes from the directory names, else from the names at the end of PATH, else from AQ_mod.

vendor (info --view full) lists the first 200 spots with every acqus key (spots: path, acqus; spot_count).

check

check --headers-only parses every acqus (no_calibration, too_many_spots past 200,000 spots). check also reads every fid and computes its m/z (short_fid when fid holds fewer points than TD, bad_spectrum).

Vocabulary

identifier meaning
BrukerFlexReader the flex FormatReader
FLEX_FORMAT_ID bruker-flex
FlexDataset an open run; spectrum returns one spectrum with its points
Params, parse, get, is_xmass, Calibration, from, mz, is_spot, find_spots, dw, delay, ml1, ml2, ml3, map parameter file and calibration (sections above)
modes, total_points, instruments, instrument_types, software_versions, spots, path, spot, mode, points, shots, instrument, instrument_type, software_version, acquired, comment, calibration, processed, mz_note, acqus, spot_count JSON keys (tables above)
no_calibration, too_many_spots, short_fid, bad_spectrum check finding codes
linear, reflector, LIFT, unknown acquisition modes