Metadata conventions and the experiment model

Each reader maps its vendor’s metadata onto one model with OME-style names. The same rules apply to every format, so a script written against a CZI file also works on an ND2 or a LIF file. This page lists those rules: units, timestamps, colours, axes and missing values. It then describes the experiment block, which says what a file’s contents mean as an experiment, and how the model maps onto OME-XML when you export.

For a first tour of the JSON output, read Reading the JSON output.

Missing values

  • Missing is omitted. If the file doesn’t record a value, it is left out of the JSON rather than written as 0, "" or a default. A value the file stores as 0 where 0 is impossible, such as the pixel size of an uncalibrated axis, also counts as missing.
  • Empty text is missing. A text field stored empty or blank is left out, just like one the file doesn’t have, so a missing key means either “not stored” or “stored empty”. Two kinds of name are always present, even when empty: table column names and signal-channel names, because an unnamed column is still a column.
  • Vendor text is kept verbatim in the vendor tree of info --view full. The normalized field holds the converted value. When a reader converts local text, it keeps the original in extra, for example ND2’s extra.acquired_at_text.
  • Text is trimmed. Leading and trailing blanks are removed from objective and instrument names; inner spacing is kept.

Units

No field carries a unit string, except physical_size.unit (always µm) and the free-text units of signal channels and table columns. The unit is fixed by the field name or by this table:

quantity unit UCUM fields
length, pixel size, stage position µm um physical_size.{x,y,z}, *_um
wavelength nm nm excitation_nm, emission_nm, emission_band_*_nm
time, interval s s time_increment_s, start_s, delta_t_s, *_s
short durations ms ms fields ending in _ms (exposure_ms)
sampling rate Hz Hz sample_rate_hz
mass-to-charge m/z {m/z} mz, *_mz (precursor_mz)
ion mobility V·s/cm² (1/K0) V.s/cm2 inverse_reduced_mobility
volume nL nL volume_nl
temperature °C Cel *_c (camera_temperature_c)

Irregularly sampled traces, such as chromatograms, have sample_rate_hz 0. They set extra.irregular_sampling, and channel 0 (time) holds each sample’s time in seconds.

Timestamps

  • Timestamps are ISO-8601 with a T: YYYY-MM-DDTHH:MM:SS[.fraction] and a zone. Readers convert to UTC and write Z. An explicit ±hh:mm offset is kept when the source writes one.
  • When the source records local time with no zone, the value has no zone designator, and the file’s notes include a sentence that says so. FCS files, Waters .raw headers and some ChemStation dates are like this.
  • Fractional seconds keep the source’s precision.
  • Years outside 1900 to 2100 come from unset clocks and give no timestamp.
  • Date-only values are YYYY-MM-DD.

Colours

channels[].color and ROI colours are #RRGGBB: six upper-case hex digits, no alpha. They record how the acquisition software displayed the channel, not what the dye emits. A white channel (#FFFFFF) is a grey-scale display.

Images and axes

  • images[].index runs from 0 and is what --image takes.
  • size_x, size_y, size_z, size_c and size_t are all at least 1. An image’s plane_count is size_z × size_c × size_t; the file’s plane_count is the sum over images.
  • dimension_order is XY followed by C, Z and T in the order the source stores its planes, fastest first, as in OME. It is for information only. Planes are always addressed by (c, z, t), and exports always write XYCZT.
  • channels[i].index is the zero-based C index, the value --select c= takes. Channels are listed in order, one per C.
  • samples_per_pixel is 1 for grey or 3 for RGB. An RGB image has one channel per C; its three samples are not separate channels.
  • pyramid_levels is at least 1; 1 means no pyramid.

Channels

Emission. emission_nm is the single emission wavelength the file records, and its meaning differs by format. When the file records a detection band, such as a spectral detector window or an emission filter’s edges, emission_band_start_nm, emission_band_end_nm and emission_band_center_nm give it. Compare bands, not emission_nm, across formats.

format emission_nm detection band
CZI the channel’s emission wavelength the detection wavelength ranges
ND2 the peak of the emission spectrum the emission filter’s edges
LIF λ scans: the start of the detection window; otherwise not set the detection window or band
OIR λ channels: the λ axis wavelength; otherwise the dye’s emission start and end wavelength
OME-TIFF, OME-Zarr the channel’s EmissionWavelength the pass band of the channel’s emission filters
Zeiss LSM not set the channel’s detection range

Acquisition mode. channels[].acquisition_mode is a readable label: the technique, then the contrast method when there is one. Every reader uses the same spelling. Files that store OME enumeration values (CZI, OME-TIFF, OME-Zarr) are converted:

stored label
LaserScanningConfocalMicroscopy + Fluorescence Laser Scanning Confocal Fluorescence
WideField + Fluorescence Widefield Fluorescence
WideField + Phase Phase Contrast
BrightField + Brightfield Brightfield
SpinningDiskConfocal Spinning Disk Confocal
MultiPhotonMicroscopy Multiphoton
TIRF, TotalInternalReflection TIRF
FluorescenceLifetime Fluorescence Lifetime
PALM, STORM, STED, SPIM as written

Other vendor wording is kept, such as Brightfield (RGB) from VSI.

Immersion. objective.immersion is Oil, Water, Air, Glycerol, Multi or Other, whatever the case in the file. dry becomes Air. Other media, such as Silicone, are kept as written.

Per-frame records

info --view full lists per-plane acquisition data under images[].extra.frames: the first 100 per image, or all with --all-frames. frame_records_total gives the count. ND2 and CZI files have them. All fields are optional except the indices:

field meaning
t, z T and Z index within the image (default 0)
c C index; absent when one record covers every channel of the frame
time_ms or delta_t_s time since the image’s acquired_at
acquired_at absolute time of the frame, ISO-8601 UTC
exposure_ms exposure of the frame
exposure_ms_per_channel exposures indexed by C
stage_x_um, stage_y_um, stage_z_um stage position when the frame was taken

Other keys belong to the format, such as ND2’s pfs_status and camera_temperature_c.

Shared extra keys

extra holds each format’s own fields. These keys mean the same thing in every format, and exports and info --view explain read them:

key contents exported as
channel_settings per channel: detector, binning, gain, exposure, filters OME Detector, DetectorSettings
stage_position_um {x, y, z} OME StageLabel
refractive_index number OME ObjectiveSettings
experimenter a user name, or name parts, e-mail and institution OME Experimenter
experiment a description, or an object with type and description OME Experiment
rois regions of interest with shape, colour and geometry OME ROI
bits_significant bits the camera actually filled –
frames per-frame records (above) OME Plane

Provenance

info --view full includes a provenance map that says how the meaning of each normalized field was learned. Keys are paths into the output, such as images[].physical_size.x, with [] meaning every element. Each entry names a source, from most to least authoritative:

  • spec: a published specification or open standard, such as OME-XML or FCS 3.1.
  • vendor-impl: the vendor’s own published implementation or documentation.
  • prior-art: the public documentation of a permissively licensed community reader.
  • inferred: worked out by comparing files in the test corpus.

Reading the JSON output explains how to use these when deciding how far to trust a value. Each format’s notes in Formats list the source of every field.

The experiment block

The experiment block says what a file’s contents mean as an experiment: which sample, on which instrument, with which method, when, by whom, and what was measured, in scientific words. It has the same shape for a confocal stack, an LC-MS run, a flow-cytometry tube and a plate read, so you can ask the same question of any file. It is left out when the file records none of this.

Terminal window
openreadout info run.raw --json | jq .data.experiment
{
"sample": {"id": "QC1", "sequence_position": "D:35", "source_field": "spectra[0].extra.sample_name"},
"instrument": {"vendor": "Thermo Fisher Scientific", "model": "LTQ Orbitrap Discovery", "serial": "20698",
"software_version": "2.4 SP1", "kind": {"id": "OBI:0000049", "label": "mass spectrometer"}},
"method": {
"name": "UPLC-ESI-profiling-19min_profile_NEG",
"technique": {"id": "CHMO:0000524", "label": "liquid chromatography-mass spectrometry"},
"assay": {"id": "OBI:0003097", "label": "liquid chromatography mass spectrometry assay"},
"parameters": {
"method_length": {"value": 19, "unit": "min", "ucum": "min"},
"injection_volume": {"value": 10, "unit": "µL", "ucum": "uL"},
"polarity": {"value": ["negative"]}, "ms_levels": {"value": [1]}, "ion_source": {"value": "ESI"}
}
},
"acquisition": {"started_at": "2009-05-07T20:03:10.046Z", "operator": "LTQ OT Discovery", "duration_s": 1139.976},
"measurements": [
{"kind": "spectra", "indices": [0],
"what": "LC-MS, negative mode, ESI, 2,048 scans, MS1, retention time 0.0119–19 min, m/z 50–1000",
"technique": {"id": "CHMO:0000524", "label": "liquid chromatography-mass spectrometry"},
"terms": [{"id": "MS:1000129", "label": "negative scan"}, {"id": "MS:1000073", "label": "electrospray ionization"},
{"id": "MS:1000579", "label": "MS1 spectrum"}, {"id": "MS:1000484", "label": "orbitrap"}],
"parameters": {"scans": {"value": 2048}, "mz_range": {"value": [50, 1000], "unit": "m/z", "ucum": "{m/z}"}}}
],
"provenance": {
"sample.id": {"source": "inferred", "from": "spectra[0].extra.sample_name"},
"instrument.model": {"source": "prior-art", "from": "spectra[0].instrument.model"},
"method.technique": {"source": "inferred", "from": "spectra[0]"},
"measurements[0]": {"source": "inferred", "from": "spectra[0]"}
}
}

The parts:

  • sample: what was measured. Fields are id, name, well, barcode and sequence_position (vial or autosampler position). source_field names the field the id came from.
  • instrument: vendor, model, serial, software, software_version, and kind, an OBI device term.
  • method: the method, protocol or experiment name as saved by the software; the technique (a CHMO, FBbi or OBI term); the assay (an OBI term); and parameters.
  • acquisition: started_at, ended_at, operator, duration_s and comment. duration_s is the length of the recorded data. comment is the free text saved with the acquisition, such as an ABF file comment or FCS $COM. saved_at appears instead of started_at when the file records only when it was saved, as SoftMax Pro text exports do.
  • measurements[]: one entry per kind of data block (image, table, trace, spectra), with indices pointing at the blocks. what describes the measurement in words, such as absorbance at 450 nm, 96 wells.
  • notes: things to know, such as a CZI whose scenes cover several wells.
  • provenance: where each value came from.

Parameters use our own snake_case names: nucleus, pulse_program, method_length, injection_volume, polarity, ms_levels, ion_source, wavelength, objective_magnification, pixel_size, time_interval, sample_rate, gradient, solvents, oven_program, column and others. Each is {value, unit, ucum}; a parameter with a unit carries its UCUM code. Structured settings keep their units in their keys:

  • gradient is the LC pump program as steps {time_min, flow_ul_min, percent: {A: …, B: …}}.
  • solvents names the solvent channels: {"A": "Eau + 0.1% HCOOH", "B": "ACN + 0.1% HCOOH"}.
  • oven_program is the GC oven program as steps {temperature_c, hold_min, rate_c_per_min}.

Provenance. Every value has an entry keyed by its path in experiment, such as sample.id or method.parameters.nucleus. An entry covers everything under its path. from is the field the value was taken from. source is the reader’s own source for that field, or inferred for OpenReadout’s own mapping, such as the choice of sample-id field, the what text, terms and durations. Values a reader takes from a vendor file name that file, such as acqus ##OWNER.

Where the sample id comes from

sample.id is the first of these the file records:

  • Mass spectrometry and chromatography: the sequence’s sample id, then its sample name, then the mzML sampleList. The vial or autosampler position becomes sequence_position.
  • Flow cytometry: $SMNO, then the vendor keywords TUBE NAME, SAMPLE ID or SampleID, then $WELLID, then $SRC. $WELLID also gives well, and $CELLS gives name.
  • Plate readers: the plate barcode or id, else the plate name when it is not a default such as Plate 1.
  • NMR: Bruker’s USERA1 to USERA5 fields, then the first line of the title; JCAMP-DX ##TITLE; JEOL’s sample_id.
  • Microscopy (CZI, ND2 and LIF only): the plate well of the scenes when all images share one, else the image name when every image has the same one and it looks like a sample id.

Software defaults are not used as sample ids: Position 1, Series011, TileScan_002_Merging, Plate 1, Specimen_001, file names, and placeholders such as NA, -1 and <user>. A LIF project name describes the experiment, not one sample, so it becomes method.name.

Terms and units

Terms come from a curated table of ontology ids and labels; no ontology files are shipped.

where vocabulary examples
instrument.kind OBI devices OBI:0400169 microscope, OBI:0400044 flow cytometer, OBI:0000049 mass spectrometer, OBI:0001058 microplate reader
method.technique CHMO (chemistry), FBbi (imaging), OBI CHMO:0000524 LC-MS, CHMO:0000593 1H NMR, CHMO:0000061 flow cytometry, FBbi:00000246 fluorescence microscopy
method.assay OBI assays OBI:0003097 LC-MS assay, OBI:0000916 flow cytometry assay, OBI:0002119 microscopy assay
measurements[].terms PSI-MS, CHMO, FBbi MS:1000129 negative scan, MS:1000073 ESI, MS:1000579 MS1 spectrum, FBbi:00000249 time lapse

Units are written for people and carry their UCUM code: µm → um, Å → Ao, µs → us, °C → Cel, µL → uL, ppm → [ppm], m/z → {m/z}, cm⁻¹ → /cm, mM → mmol/L, rpm → {rpm}. Plate-reader units are UCUM annotations: OD → {OD}, RFU → {RFU}, RLU → {RLU}.

ontology version checked licence source
PSI-MS controlled vocabulary (MS:) 4.2.2 CC BY 3.0 https://github.com/HUPO-PSI/psi-ms-CV
Chemical Methods Ontology (CHMO:) 2026-05-28 CC BY 4.0 https://github.com/rsc-ontologies/rsc-cmo
Biological Imaging Methods Ontology (FBbi:) 2026-06-25 CC BY 4.0 https://github.com/CRBS/Biological_Imaging_Methods_Ontology
Ontology for Biomedical Investigations (OBI:) 2026-07-27 CC BY 4.0 https://github.com/obi-ontology/obi
Unified Code for Units of Measure (UCUM) 2.2 UCUM License (no charge, royalty-free) https://ucum.org/license

We checked each term id and label against the EBI Ontology Lookup Service for the release listed. Attribution is in the repository’s NOTICE file.

Asking in words

info --view explain opens with an Experiment paragraph: sample, technique, method and settings, operator and data length.

info FILE --ask "QUESTION" matches the question’s words to topics and answers each from the model, naming the fields each answer came from:

Terminal window
openreadout info run.raw --ask "what was the gradient?"
topic words that select it
sample sample, specimen, well, barcode, vial, tube, position
channels channel, dye, stain, DAPI, GFP, wavelength, excitation, emission, filter, detector
method gradient, oven, solvent, column, method, protocol, pulse program, settings, parameter
run_length how long, duration, run length, run time
polarity polarity, positive, negative, ion mode
ms_levels MS level, MS/MS, MS2, tandem, fragment, precursor
comment comment, note, remark, annotation
instrument, operator, acquired, technique instrument or model; who or operator; when or date; technique or modality

A question that matches no topic gets the technique, sample, method and run length.

MS1 and MS/MS scans are interleaved in a run, so “the first MS/MS scan” is not simply the first scan after the MS1 scans. openreadout spectra FILE --ms-level 2 --nth 1 returns it directly.

OME-XML export

export --to ome-tiff writes OME-XML, as does the OME/METADATA.ome.xml of a multi-image OME-Zarr store. It validates against the OME 2016-06 schema. Lengths avoid non-ASCII characters so that Bio-Formats can read them: PhysicalSizeX/Y/Z carry no unit attribute (the schema default is µm), and stage and plane positions are written in nm.

OME element from
Experiment extra.experiment
Experimenter extra.experimenter
Instrument/Microscope instrument.manufacturer, instrument.model
Instrument/Laser a channel’s excitation_nm when its acquisition mode names a laser technique (confocal, multiphoton, TIRF, STED)
Instrument/GenericExcitationSource other channels’ excitation_nm
Instrument/Detector instrument.detector, then each other detector in extra.channel_settings
Instrument/Objective objective
Instrument/Filter each distinct detection band
Image/@Name name, when the source has one
Image/AcquisitionDate acquired_at, to the millisecond
Image/ObjectiveSettings objective.immersion, extra.refractive_index
Image/StageLabel extra.stage_position_um, else the first frame’s stage position
Pixels/@PhysicalSize*, @TimeIncrement physical_size, time_increment_s, scaled by the step of an even --select
Channel name, fluor, wavelengths, colour, acquisition mode and contrast method
Channel/DetectorSettings gain and binning from extra.channel_settings
Plane per-frame records; DeltaT falls back to time_increment_s × t
ROI extra.rois; coordinates are copied as stored and are not verified
MapAnnotation (namespace openreadout.dev/normalized) what OME cannot hold exactly: the full-precision acquired_at, the acquisition software and version, an immersion outside the OME list, an acquisition-mode label OME cannot express

EmissionWavelength is written only when the file records an emission wavelength; a band is not turned into one. Images with identical instruments, objectives, light sources, detectors and bands share one Instrument. --embed-vendor also stores the vendor metadata tree.

OpenReadout’s own OME-TIFF and OME-Zarr readers read the MapAnnotation back, so a round trip keeps those values. Other OME readers see ordinary key-value annotations.

An export and read-back does not reproduce:

  • dimension_order, which is always XYCZT in the export.
  • The source’s tiling (mosaic, tile sizes). The export holds the stitched image in its own tiles.
  • time_increment_s where the source had only per-frame times. The OME-TIFF reader infers the mean step from the planes, so the read-back has one.
  • Everything in extra except what the table above maps.

check --against compares a source with its export and reports these differences; see check.

Conformance

The corpus test crates/openreadout-corpus-tests/tests/metadata.rs checks every test file against the rules on this page. For the experiment block it checks that every value has a provenance entry, every term is in the curated table, every unit has its UCUM code, timestamps are ISO-8601, duration_s is positive and every measurement has a description. Maintainers run it as described in docs/maintaining.md.