Agilent ChemStation saves each HPLC or GC acquisition as a .D directory. OpenReadout returns the detector signals as traces, diode-array data as one trace per file with a channel per wavelength, single-quadrupole MS data as spectra, and the vendor’s peak report as a table.

Derived from hex dumps of public corpus data sets, then checked against permissively licensed prior art (entab, MIT; Aston, BSD-3-Clause) and rainbow’s public documentation pages; rainbow (LGPL-3.0) and Aston are used as reference readers. See docs/provenance/chemstation.md. Format id: chemstation; crate openreadout-chrom (chemstation_*.rs).

A ChemStation acquisition is a directory ending in .D (or .d). Each detector signal is a file of its own: .ch (one channel over time: FID, TCD, VWD, MWD/DAD single wavelengths, ADC), .uv (diode-array spectra over time), .ms (single-quadrupole mass spectra). The directory also holds .REG register dumps, RUN.LOG, SAMPLE.XML/SAMPLE.MAC, acqmeth.txt, .ini files and the method directory *.M/. OpenReadout opens either the directory or a single signal file.

We expose one trace per .ch file (one channel), one trace per .uv file (one channel per wavelength), and one spectra run per .ms file. Every other file is listed by info --view structure with its size and version string; text files (.txt, .xml, .ini, .log, .mac, .mth, .csv) are copied into the info --view full vendor tree (first 16 KiB).

Version string (all files)

Byte 0 is a length (1–3), followed by that many ASCII digits: the file’s version. Seen: 2 (.ms), 30, 81, 130, 179, 181 (.ch), 131 (.uv); register (.REG) files start with a version string as well. Detection (looks_like_chemstation): versions below 100 must have a length-prefixed type string ending in DATA FILE (GC DATA FILE, LC DATA FILE), or MSD Spectral File/GC / MS Data File, at 0x04 (or a zero byte after the digits, for register files); versions ≥ 100 repeat the version as UTF-16 at 0x146. A .d directory is ours when at least one top-level .ch/.uv/.ms file passes this test (Agilent MassHunter .d directories, which hold AcqData/, do not).

The header length is a big-endian u32 at 0x108 counted in 512-byte blocks plus one (3 → 1,024 bytes, 13 → 6,144 bytes; (n − 1) × 512). For .ms files it is a big-endian u16 at 0x10A in 16-bit words plus one (257 → 512 bytes).

Strings

Versions below 100: one length byte, then single-byte characters (read as Latin-1, trimmed). Versions 130/131/179/181: one length byte (in characters), then UTF-16LE code units.

our name versions < 100 130/131/179/181 example
file_type 0x04 0x15B GC DATA FILE, LC DATA FILE, MSD Spectral File
sample_name 0x18 0x35A Cytochrome C, BB7125_3-spiropyrrolidine_cof
description 0x56 – (spaces in the corpus)
operator 0x94 0x758 RJB, SYSTEM
acquired_text 0xB2 0x957 18-Nov-10, 15:48:06, 8/20/20 2:00:32 PM, 28 Jun 13 10:59 am -0500
instrument_model 0xD0 0x9BC HP G1530A, G1365B, GCI (a generic GC module tag, not a model), DAD1
separation 0xDA 0x9E5 GC, LC
method 0xE4 0xA0E EVAL.M, RJB-TEST.M
instrument_name – 0xC11 Asterix ChemStation
software / software_revision – 0xE11 / 0xEDA A.01.15 / Rev. C.01.07 SR2 [2...
units 0x244 0x104C (0xC15 in .uv) pA, mAU, 25 uV
signal 0x254 (0x140 in .ms) 0x1075 FID1A, Front Signal, MWD A, Sig=210,5 Ref=360,100, MSD1, Initial Scan Range=100.0-1000.0
vial_text – 0xFD7 5 (equals the u16 at 0xFE)

Every other length-prefixed string in the header is found by a scan (a length byte after a zero byte, followed by printable characters and a zero) and listed in info --view full under its hex offset. parse_date turns acquired_text into ISO-8601 (acquired_at); two-digit years < 70 are 20xx; a trailing ±hhmm becomes the zone.

wavelengths reads the detector wavelength and bandwidth from Sig=210,5 and the reference from Ref=360,100 (Ref=off gives none) → trace extra.wavelength_nm, bandwidth_nm, reference_wavelength_nm, reference_bandwidth_nm.

Numbers (big-endian)

offset our name type notes
0xF8 – u32 the version again as a number (81, 179, 181)
0xFC / 0xFE / 0x100 sequence_line / vial / replicate u16 names from entab (prior art); vial corroborated by the 0xFD7 string
0x116 declared_records u32 scan count in .uv files and LC-MS .ms files (GC-MS .ms: little-endian u16 at 0x142); not a count in .ch files
0x11A / 0x11E first_time_ms / last_time_ms f32 (81, 179, 181), i32 (30, 130) retention time in milliseconds
0x27C / 0x284 offset / scale f64 versions 30/81: value = raw × scale + offset (offset 0 in every corpus file)
0x127C scale f64 versions 130/179/181
0xC0D scale f64 version 131 (.uv)

Scales seen: 1/7,680 (FID pA), 1,000/2²¹ (MWD/DAD mAU).

Bodies (BodyEncoding)

encoding versions layout
DeltaRecords 30, 130 records: tag byte 0x10, count byte n, then n big-endian i16 first differences; 0x8000 is followed by a big-endian i32 absolute value. The body ends with two zero bytes (Terminator).
SecondDifference 81, 181 big-endian i16 second differences (d += w; value += d); 0x7FFF is followed by a 48-bit big-endian absolute value (i32 high word, u16 low word: high × 65,536 + low) and resets d to 0. Version 181 ends with one 0x0000 word, which is a terminator, not a sample.
Float64 179 little-endian f64 values to the end of the file
SpectrumRecords 131 per scan: u16 tag 67, u16 record length, u32 time (ms), u16 first/last/step wavelength × 20, 8 unknown bytes, then first differences (little-endian i16, 0x8000 → i32 absolute), starting from 0 in each record; or (OpenLab CDS spectra parts) tag 70 with little-endian f64 values after the same 22-byte header, value = f64 × scale × scale
MassSpectrumRecords 2 per scan: u16 length in 16-bit words, u32 time (ms), 6 bytes, u16 pair count, 4 bytes, pairs of u16 m/z × 20 and u16 packed intensity (14-bit mantissa × 8^(top two bits)), then 10 bytes whose last u32 is the scan’s TIC; all big-endian
NotDecoded others listed only

BodyEnd records how a walk ended: Terminator, EndOfFile, Truncated { at } or BadRecord { at, detail }.

Time axis

Traces are regular: start_s = first time; the interval is (last − first) / (n − 1) for versions 30/130/131/179/181 and (last − first) / n for version 81 (both corpus version-81 files span exactly n intervals of 200 ms or 100 ms). For .uv files the times come from the records. extra.axis = {quantity: retention_time, unit: min, first, step}; extra.x_start_min / x_end_min are the header times in minutes.

Mapping

  • traces[]: .ch → name = the signal string (else the file stem); one channel named after the file stem (FID1A, mwd1A), unit from the header, dtype = raw storage (int32 for delta records, int64 for second differences, float64), scale/offset from the header. .uv → name <stem> spectra, one channel per wavelength (200 nm, …, extra.wavelength_nm), extra.wavelength_range_nm = [first, last, step]; records on another wavelength grid are placed by wavelength (NaN elsewhere) and extra.wavelength_grid_varies is set.
  • extra (both): file, format_version, encoding, file_type, sample_name, description, method, operator, acquired_at, acquired_text, instrument, instrument_name, separation, software, software_revision, signal, vial, detector (letters that start the file stem’s last part after -, _, . or a space: FID, MWD, DAD, VWD, TCD), result_modules (from Result.xml), wavelengths, x_start_min, x_end_min, axis.
  • extra.method_file (both, .D directories with an acqmeth.txt): the method report ChemStation writes, parsed by parse_acqmeth (§ Acquisition method text).

Acquisition method text (acqmeth.txt)

Latin-1 text with CRLF lines (read up to MAX_METHOD_TEXT, 1 MiB). Only mtbls75-x-fsfa-hl-gc-o7c-1-d (GC-MSD) has one in the corpus; no LC .D does, so no pump timetable is parsed. Recognized lines, anywhere in the text:

text method_file key
INSTRUMENT CONTROL PARAMETERS: GCMSD_1 instrument_name
a line holding only a path ending in .M (C:\MSDCHEM\1\METHODS\PNNL_METABOLOMICS.M) method_path
Oven Program On, then 60 °C for 1 min, then 10 °C/min to 325 °C for 10 min, … oven_program[] {temperature_c, hold_min, rate_c_per_min (ramps)}
the Run Time 37.5 min line right after the oven program run_time_min
Injection Volume 1 µL injection_volume_ul
Mode Splitless right after an … Inlet … heading inlet_mode
Column #1, its name line, 325 °C: 30 m x 250 µm x 0.25 µm column (HP-5MS 5% Phenyl Methyl Silox, 30 m x 250 µm x 0.25 µm)
Solvent Delay : 6.50 min, Low Mass : 50.0, High Mass : 600.0, Acquistion Mode : Scan (sic) solvent_delay_min, low_mass, high_mass, ms_acquisition_mode
TUNE PARAMETERS for SN: US92032548 ms_serial (the MSD’s serial number)

The experiment model maps oven_program, run_time_min (as method_length), injection_volume_ul, column, inlet_mode and ms_serial (instrument.serial). The instrument model skips names that are configured instrument names, not models: the default Instrument 1 (cut to Instrumen in .MS headers, entab-carotenoid-extract-d) and module instances named like their signal files (DAD1, zhulong-001-1-sm-d); the next candidate (the .uv header’s G1315B) is used, or none. The first Result.xml module’s name (Agilent 7890A) comes before the signal headers’ string and its serial number is instrument.serial; the generic GC tag GCI is never a model.

  • spectra[]: .ms → scan_count, rt_range_s, ms_levels [1], extra.max_pairs_per_scan; read_spectrum returns the pairs in ascending m/z (files store them descending), total_ion_current from the record trailer, polarity unknown, centroided false.

Vendor peak reports (tables[0] vendor_peaks, 2026-09-26)

Read from the top level of a .D directory (at most MAX_REPORT_BYTES, 4 MiB each):

  • Result.xml (LC/GC ChemStation’s XML export, UTF-16; preferred over Report.TXT, which prints the same integration): root ChemStationResult; Chromatograms/Signal (Description FID1 A, Front Signal → signal FID1 A, YUnits → area unit pA*s, height unit pA) with one IntegrationResults per peak (RetTime, Area, AreaPercent, Height, Width, Symmetry, TimeStart, TimeEnd, BaselineStart, BaselineEnd: the integration limits and the vendor’s baseline value at each), and Results/ResultsGroup/Peak compound results (SignalDesc, MeasRetTime, PeakType, Name, Amount with its Unit, CompoundID), joined to the integration result of the same signal and retention time (a compound result with none is counted in rows_skipped). CalibrationInformation (levels, curves) stays in the file.
  • Report.TXT (LC/GC ChemStation’s printed report; UTF-16LE with a BOM, or 8-bit): after each Signal <n>: <detector> <letter>, … line, a table of two header lines, a separator of dash runs each closed by |, and rows up to Totals : or a blank line. The separator gives the column extents (the | belongs to the column on its left); the header lines cut at the same extents give each column’s name and its unit in brackets. Columns used by name: Peak #, RetTime [min], Type, Width [min], Area [unit], Height [unit], Area %, Amount, Name; others stay in extra.reports[].columns. The signal FID1 A names the trace of FID1A.ch (spaces removed, case ignored).
  • RESULTS.CSV (MSD ChemStation): INI-like; each [INT <signal>] section has Header=,"Peak","R.T.","First","Max","Last","PK TY","Height","Area","Pct Max","Pct Total" and <n>=, rows. First/Max/Last are 1-based scan numbers. INT TIC: <run>\data.ms names the TIC of that .ms file (TIC data.ms).

Columns: signal (code into extra.categories), peak, rt_min, peak_type (code: BB, BB S, M2, …), width_min, area, height (unit when every table states the same one: pA*s / pA), area_pct, amount (its unit when all amounts share one, e.g. % v/v), compound (code), first_scan, max_scan, last_scan, start_min, end_min, baseline_start, baseline_end, symmetry (NaN where the report has no such column). extra.source vendor; extra.reports[]: file, signal, signal_text, area_unit, height_unit, columns (as written), peaks, rows_skipped (rows that did not parse as peaks).

check finds every reported peak above 1 % of its table’s largest on our decoded signal: our maximum within half the reported width (at least one sample) of the retention time, not at the window edge, and our height within 10 % of the reported one — above the vendor’s baseline when Result.xml gives it, else above a line through the trace 3 widths each side; with Result.xml also the area, our trace integrated between the vendor’s limits above its baseline, within 10 %; RESULTS.CSV — the apex scan’s time within one scan of the retention time and its TIC within 10 % of the reported height. Peaks whose type carries a flag after its two baseline letters (BV E, VV R: skimmed or special baselines) are not compared. When more than a quarter of the compared peaks differ (a wrong scale, unit or signal moves them all) check warns vendor_peak_mismatch; fewer are reported as vendor_peak_outliers / vendor_area_outliers (info: a neighbour inside the window in a crowded region). The MSD Area is the integrator’s corrected area (about 3.6× our sum of TIC counts, depending on the peak type) and is not compared.

Checked on the corpus (tests/chemstation_reports.rs): 13 Report.TXT peaks (4 GC runs, FID and TCD), 131 Result.xml peaks (2 GC runs, FID and TCD; 22 named compounds with amounts) and 46 RESULTS.CSV peaks (12 GC-MSD runs) read and found on our signals — the Result.xml areas recomputed between the vendor’s limits agree to a median ratio of 0.9999 (66 peaks above 1 % of the largest area; the few that differ are skimmed or in a crowded region and are reported as info); the runs’ tic_front.csv TIC exports equal our TIC at every one of 24,144 scans.

check finding codes

bad_header_length, truncated, bad_record (errors); unknown_version, no_samples, bad_time_range, scan_count_mismatch, time_not_monotonic, vendor_peak_mismatch (warnings); no_terminator, time_range_mismatch, wavelength_grid_changes, vendor_report_rows_skipped, vendor_peak_outliers, vendor_area_outliers (info).

Vocabulary (every public identifier in chemstation_*.rs must appear here)

identifier meaning
ChemStationReader, ChemStationDataset, CHEMSTATION_ID format reader, opened .D directory or file, the id chemstation
open, signals, files, is_chemstation_dir, count open a directory/file; its decodable signal files; every file; directory test; samples or scans in a body
DirFile, rel, size, version one file of the directory: relative path, bytes, version string
SignalFile, path, header one decodable file: absolute path and parsed header
MAX_METHOD_TEXT, parse_acqmeth, method_file acqmeth.txt method report
ReportPeak, number, rt_min, peak_type, width_min, area, height, area_pct, amount, first_scan, max_scan, last_scan one peak of a vendor report (§ Vendor peak reports)
ReportTable, source, area_unit, height_unit, columns, peaks, skipped, reports one signal’s table of a report (Report.TXT / RESULTS.CSV); a data set’s reports
ResultModule, name, serial, firmware, part_number, parse_result_modules, result_modules one instrument module of Result.xml (ModuleInformation/Module: ModuleName, SerialNumber, FirmwareRevision, PartNumber); given as extra.result_modules
parse_report_txt, parse_results_csv, parse_result_xml, decode_text, MAX_REPORT_BYTES report parsers; UTF-16 (BOM) or 8-bit text; largest report read
start_min, end_min, baseline_start, baseline_end, symmetry, amount_unit, compound_id, blank, flagged Result.xml peak fields; an empty peak; a peak type with a flag (skimmed/special baseline)
SignalHeader, kind, encoding, header_len, file_type, sample_name, description, operator, acquired_text, acquired_at, instrument_model, separation, method, instrument_name, software, software_revision, units, signal, vial, sequence_line, replicate, first_time_ms, last_time_ms, scale, offset, declared_records, strings header fields (tables above)
parse, wavelengths, to_json, name parse a header; Sig=/Ref= wavelengths; header as JSON; an encoding’s JSON name
version_of, looks_like_chemstation, parse_date, DECODED_VERSIONS version string at byte 0; detection; ChemStation dates to ISO-8601; versions whose body is decoded
SignalKind { Chromatogram, Spectra, MassSpectra, Other } what a file holds
BodyEncoding { DeltaRecords, SecondDifference, Float64, SpectrumRecords, MassSpectrumRecords, NotDecoded } body layouts (table above)
BodyEnd { Terminator, EndOfFile, Truncated, BadRecord }, at, detail how a body walk ended
DecodedSignal, values, escapes, records, end decoded .ch body: raw values, absolute-value escapes, delta records walked, how it ended
decode_delta_records, decode_second_difference, decode_float64 .ch body decoders
ScanRecord, len, time_ms, wavelength_raw, tic, float a located .uv/.ms record: file offset, bytes, time, wavelength range × 20, values or pairs, stored TIC, f64 values (tag 70)
UV_TAG_DIFFS, UV_TAG_FLOAT .uv record tags: 67 (16-bit differences), 70 (f64 values, scaled by the header scale twice; OpenLab CDS spectra parts)
UV_RECORD_HEADER, MS_RECORD_HEADER, MS_RECORD_TRAILER record header/trailer sizes (22, 18, 10 bytes)
uv_record, uv_values, ms_record, ms_pairs, ms_intensity record header parsers and value decoders; packed MS intensity

How this reader was derived, file by file: provenance log.